# Name: M_TLIMITON # # Date: # # created: 08 15 1994 SDW # revised: # # Change Description # # None # # MDI Author: Scott Williams 415-725-0482 # # Description # # This procedure set limits for the MDI internal temperatures during the # warm up period following a power on. # # Intended Use # # Establish initial limits for MDI internal temperature monitors during # the rise to stable operatiional temperature. Excedence of these # limits may indicate hardware risk. # # Duration # # ? # # Requirements and Constraints # # MDI in SLEEP or higher mode. This procedure should be completed # as part of the MDI power on. All temperatures TBR. # # Steps # # Step Description Command Monitor Value # 0 set temp limits for CCD strap # limit qtmi1dn n/a n/a 40.0 50.0 qtmi1dn # 1 set temp limits for Optics Package # limit qtmi1bn 5.0 10.0 30.0 35.0 qtmi1bn # 2 set temp limits for Oven # limit mtopts4 25.6 25.6 36.0 37.0 mtopts4 # limit mtopts5 25.6 25.6 36.0 37.0 mtopts5 # 3 end of procedure